ASTM E1127-03

system ASTM 2017-08-16 9 PDF 413 0 2026-07-04 01:28
summary
Standard Guide for Depth Profiling in Auger Electron Spectroscopy
description

Significance and Use

Auger electron spectroscopy yields information concerning the chemical and physical state of a solid surface in the near surface region. Nondestructive depth profiling is limited to this near surface region. Techniques for measuring the crater depths and film thicknesses are given in (35).

Ion sputtering is primarily used for depths of less than the order of 1 μm.

Angle lapping or mechanical cratering is primarily used for depths greater than the order of 1 μm.

The choice of depth profiling methods for investigating an interface depends on surface roughness, interface roughness, and film thickness (1).3

Scope

1.1 This guide covers procedures used for depth profiling in Auger electron spectroscopy.

1.2 Guidelines are given for depth profiling by the following:

Section
Ion Sputtering6
Angle Lapping and Cross-Sectioning7
Mechanical Cratering8
Nondestructive Depth Profiling9

1.3 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Details
Publisher ASTM
File size 70.83 KB
Pages 5
Format PDF
Publication date 2017-08-16