summary
Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants
description
Significance and Use
This test method provides a quantified measure of the image artifact produced under a standard set of scanning conditions.
This test method applies only to passive implants that have been established to be MR-Safe or MR-Conditional.
Scope
1.1 This test method characterizes the distortion and signal loss artifacts produced in a magnetic resonance (MR) image by a passive implant (implant that functions without the supply of electrical or external power). Anything not established to be MR-Safe or MR-Conditional is excluded.
Details
Publisher
ASTM
File size
55.13 KB
Pages
4
Format
PDF
Publication date
2013-03-11