NOTE 1 It is possible to use other ranges depending on the levels present in ambient air.
NOTE 2 Exemplar uncertainty budget calculations are given in Annexes F to H referring to Directive 2008/50/EC[1]. In the event that the Limit Values are updated in future iterations of Directive 2008/50/EC[1], the user can use these new values to calculate measurement uncertainties.
The method covers the determination of ambient air concentrations of nitrogen dioxide and nitrogen monoxide in zones classified as rural areas, urban-background areas, traffic-orientated locations and locations influenced by industrial sources. The results are expressed in µg/m3 (at 20 °C and 101,3 kPa).NOTE 3 500 µg/m3 of nitrogen dioxide corresponds to 261 nmol/mol of nitrogen dioxide at 20 °C and 101,3 kPa. 1 200 µg/m3 of nitrogen monoxide corresponds to 962 nmol/mol of nitrogen monoxide at 20 °C and 101,3 kPa.
This document contains information for different groups of users. Clause 5 to Clause 7 and Annex B and Annex C contain general information about the principles of NOx measurement by chemiluminescence analyser and sampling equipment. Clause 8, Annex E is specifically directed towards test houses and laboratories that perform type testing of NOxanalysers. These sections contain information about:- — type testing conditions, test procedures and test requirements;
- — analyser performance requirements;
- — evaluation of the type testing results;
- — evaluation of the associated uncertainty of the measurement performed by the NOxanalyser based on the type testing results.
- — initial installation of the analyser in the monitoring network and acceptance testing;
- — ongoing quality assurance/quality control;
- — calculation and reporting of measurement results;
- — evaluation of the associated uncertainty of the measurements under practical monitoring conditions.
NOTE 4 Analysers type tested prior to the publication of this document can still be used for regulated monitoring purposes. As newer versions of analysers tested under this document become available, discontinue the use of older reference analysers.