NOTE 1 It is possible to use other ranges depending on the levels present in ambient air.
NOTE 2 Exemplar uncertainty budget calculations are given in Annexes E to H referring to Directive 2008/50/EC[1]. In the event that the Limit Values are updated in future iterations of Directive 2008/50/EC[1], the user can use these new values to calculate measurement uncertainties.
The method covers the determination of ambient air concentrations of sulfur dioxide in zones classified as rural areas, urban-background areas, traffic-oriented locations and locations influenced by industrial sources. The results are expressed in µg/m3 (at 20 °C and 101,3 kPa).NOTE 3 1 000 µg/m3 of SO2 corresponds to 376 nmol/mol of SO2.
This document contains information for different groups of users. Clause 5 to Clause 7 and Annex C and Annex D contain general information about the principles of sulfur dioxide measurement by ultraviolet fluorescence analyser and sampling equipment. Clause 8 and Annex E are specifically directed towards test houses and laboratories that perform type testing of sulfur dioxide analysers. These sections contain information about:- — type testing conditions, test procedures and test requirements;
- — analyser performance requirements;
- — evaluation of the type testing results;
- — evaluation of the associated uncertainty of the measurement performed by the sulfur dioxide analyser based on the type testing results.
- — initial installation of the analyser in the monitoring network and acceptance testing;
- — ongoing quality assurance/quality control;
- — calculation and reporting of measurement results;
- — evaluation of the uncertainty of the measurement results under practical monitoring conditions.
NOTE 4 Analysers type tested prior to the publication of this document can still be used for regulated monitoring purposes. As newer versions of analysers tested under this document become available, discontinue the use of older reference analysers.