201.1 Scope, object and related standards
Clause 1 of the general standard1) applies, except as follows:
201.1.1 Scope
Replacement: This international standard applies to the basic safety and essential performance of mammographic X-ray equipment including equipment for mammographic tomosynthesis and mammographic stereotactic devices, hereafter also referred to as me equipment.NOTE 1 This includes mammographic X-ray equipment using integrated digital X-ray image receptors or integrated storage phosphor subsystems.
Excluded from the scope of this document are:- — reconstructive tomography other than mammographic tomosynthesis;
- — ct scanners covered by IEC 60601-2-44;
- — diagnostic consoles;
- — picture archiving and communication systems (PACS);
- — non-integrated storage phosphor readers;
- — hard copy cameras;
- — films, screens and cassettes;
- — computer aided detection (CAD);
- — devices for performing core biopsy and other biopsy instruments;
- — modes of operation intended to demonstrate local contrast medium uptake (contrast enhanced digital mammography);
NOTE 2 IEC 60601-2-7:1998 and IEC 60601-2-32 are not part of the 3rd edition scheme for mammographic X-ray equipment and mammographic stereotactic devices.