1 Scope and object
This part of IEC 61223 applies to those components of CTscanners which influence the image quality, patient dose and positioning.
This standard
- — defines the essential parameters which describe the performance of the CTscanners with regard to image quality, patient dose and positioning; the list of parameters to be tested can be found in section 4.4.
- — defines the methods of testing the essential parameters;
- — evaluates compliance with the tolerances of the parameters specified by the accompanying documents.
- — aspects of mechanical and electrical safety;
- — aspects of mechanical, electrical and software performance, unless they are essential for performing the acceptance tests and are directly affecting image quality, patient dose and positioning.