Passive neutron dosimetry systems. Performance and test requirements for personal dosimetry | Status: Current
description
1 Scope
This document provides performance and test requirements for determining the acceptability of neutron dosimetry systems to be used for the measurement of personal dose equivalent, Hp(10), for neutrons ranging in energy from thermal to 20 MeV1.
This document applies to all passive neutron detectors that can be used within a personal dosemeter in part or in all of the above-mentioned neutron energy range. No distinction between the different techniques available in the marketplace is made in the description of the tests. Only generic distinctions, for instance, as disposable or reusable dosemeters, are considered.
This document describes type tests only. Type tests are made to assess the basic characteristics of the dosimetry systems and are often ensured by recognized national laboratories
This document does not present performance tests for characterizing the degradation induced by the following:
- — intrinsic temporal variability of the quality of the dosemeter supplied by the manufacturer;
- — intrinsic temporal variability of preparation treatments (before irradiation and/or before reading), if existing;
- — intrinsic temporal variability of reading process;
- — degradation due to environmental effects on the preparation treatments, if existing;
- — degradation due to environmental effects on the reading process.