summary
X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy. Radiation characteristics and production methods | Status: Superseded, Withdrawn
description
1 Scope
This part of ISO 4037 specifies the characteristics and production methods of X and gamma reference radiation for calibrating protection-level dosemeters and rate dosemeters at Alr kerma rates from 10 μGy·h−1 to 10 Gy·h−1 and for determining their response as a function of photon energy. The methods for producing a group of reference radiations for a particular photon-energy range are described in four sections which define the characteristics of these radiations. The four groups of reference radiation are:
- a) in the energy range from about 7 keV to 250 keV, continuous filtered X radiation and the gamma radiation of americium-241 ;
- b) in the energy range 8 keV to 100 keV, fluorescence X radiation;
- c) in the energy range 600 keV to 1,3 MeV, gamma radiation emitted by radionuclides;
- d) in the energy range 4 MeV to 9 MeV, gamma radiation produced by reactors and accelerators.