loading preview...
summary
Biological evaluation of medical devices. Requirements for interlaboratory studies to demonstrate the applicability of validated in vitro methods to assess the skin sensitization of medical devices. | Status: Current
description
1 Scope
This document specifies the framework and the methodology to evaluate and demonstrate the applicability of a validated non-animal method from an OECD test guideline to assess the skin sensitizing potential of a medical device or a medical device material. This document addresses:
- — the database of reference chemical skin sensitizers and non-skin sensitizers;
- — reference materials;
- — feasibility testing of candidate test methods, including any method optimization for use with extracts of medical devices;
- — prevalidation of candidate test methods;
-
— the interlaboratory study:
- — sample preparation and coding;
- — spiking of the extracts from the negative control medical device material;
- — data collection;
- — statistical analysis to assess reliability and reproducibility.
Details
Publisher
BS
File size
2.26 MB
Pages
44
Format
PDF
Publication date
2023-08-02
ISBN-13
978-0-539-23352-0
ISBN-13 plain
9780539233520
ISBN-10
0539233520